PRACTICAL 2

Determination of sodium and potassium ion concentration: Methods


2.3 Experiment A: Determination of [Na+] and [K+] in solution by flame photometry

Reagents:

oral rehydration sachet

NaCl standards: 0.25, 0.5, 1.0, 2.0, 4.0 and 5.0 mM

KCl standards: 0.1, 0.2, 0.5, 1.0, 1.5, 2.0 mM

Procedure:

  1. Carefully open an oral rehydration sachet and empty the contents into a clean 250 ml beaker. Add about 150 ml distilled water and gently swirl the contents until dissolved.
  2. Pour the solution into a 200 ml volumetric flask and rinse out the beaker with small amounts of distilled water, adding the washings to the flask. Finally, make up the flask to exactly 200 ml and mix thoroughly.
  3. Make a 1/50 dilution of the redissolved sachet solution by accurately pipetting 2 ml of the solution into a 100 ml volumetric flask and making up to 100 ml with distilled water. Retain the undiluted solution for Experiment B.
  4. Instructions for use of the flame photometer:

  5. Ensure that the photometer drain is leading into a sink and that the instrument is connected to gas, air and electricity supplies. Ensure the mains supply gas tap is off.
  6. Turn the "Sensitivity" and instrument "Gas" controls control fully counterclockwise (towards you).
  7. Insert the sodium optical filter.
  8. Switch on the instrument and unclamp the galvanometer by turning counterclockwise.
  9. Open the mica window, turn on the mains gas supply, light the gas and close the window.
  10. (CAUTION: DO NOT LEAN OVER THE INTRUMENT OR YOU WILL SET YOUR HAIR ALIGHT)

  11. Turn on the air supply control and adjust the air pressure to 10 lb/in2. Leave for 1-2 minutes to stabilise.
  12. Place a beaker of distilled water into position at the left hand side of the instrument and insert the narrow draw tube into it to allow water to pass through the photometer. (NOTE: once set up, the photometer must have water running through it at all times when a salt solution is not being measured. The rate of uptake is fast, so make sure there is always enough water in the beaker).
  13. Adjust the gas control to give a flame with a large central blue cone then, with water passing through the instrument, slowly close the gas control until ten separate blue cones just form.
  14. Set the galvanometer to zero using the "Set zero" control.
  15. Replace the distilled water with the 5 mM NaCl standard and adjust the "Sensitivity" control till the galvanometer reads 100.
  16. Quickly but carefully, replace the 5 mM NaCl standard with standards of decreasing concentration from 4 mM to 0.25 mM and note the readings in the Table below.
  17. Run water through the instrument again for 1-2 min then place the draw tube into a beaker containing the 1 in 50 diluted rehydration sachet solution and note the galvanometer reading.
  18. Run water through the instrument again and replace the sodium with the potassium filter.
  19. Repeat the above procedure with the KCl standards, setting to 100 with 2.0 mM KCl, then reading the others in reverse order. Then read the 1 in 50 diluted rehydration sachet solution.
  20. Finally, run water through the instrument until the flame appears free of colour again.
  21. When the instrument is no longer required, switch off in the following sequence:
  1. Turn off the gas control and the mains gas supply
  2. Wait for the flame to die out.
  3. Turn off the air supply.
  4. Switch off the electricity
  5. Clamp the galvanometer.

[Na+] (mM)

5.0

4.0

2.0

1.0

0.5

0.25

0

Galvo reading.

100

         

0

[K+] (mM)

2.0

1.5

1.0

0.5

0.2

0.1

0

Galvo. reading

100

         

0

Plot the galvanometer readings against Na+ and K+ concentrations on the graph paper provided (separate graph for each ion) and from these calibration curves determine the Na+ and K+ concentrations in the diluted sachet solution. Finally, calculate the Na+ and K+ concentrations in the undiluted sachet solution.

 

Galvanometer reading

Diluted concentration (mM)

Undiluted concentration (mM)

Sodium ion

     

Potassium ion

     


2.4 Experiment B: Determination of [Na+] in solution with an ion-selective electrode

Reagents:

oral rehydration sachet

NaCl standards: 10, 20, 40, and 80 mM

Technical note : In order to measure the change in potential difference across the ion-selective membrane as the ionic concentration changes, it is necessary to include in the circuit a reference electrode which acts as a half-cell from which to measure the relative deviations.

Procedure:

NOTE: the bottom surfaces of the sodium ISE and reference electrode are very fragile. Do not let them drop on to the bottom of a beaker or bottle.

  1. Ensure that the multimeter (yellow box) and electrode amplifier (grey box) are connected properly with the black cable in the lowest socket (COM) and the red cable in the middle socket (VWmA) and that the silver BNC connector from the electrode head is plugged into the back of the amplifier.
  2. With the electrodes immersed to a depth of about 3-4 cm in NaCl storage solution, switch on the multimeter to the 2000 mV position (4 clicks counterclockwise) and the amplifier to position I (down). Wait until the reading has stabilised (up to 5 min.) then remove the electrodes from the storage solution, wipe the bodies (but NOT the bottoms) with a tissue, transfer to a beaker of distilled water and rinse with swirling for about 30 sec.
  3. Remove the electrodes from the water, wipe as before, then transfer to the 10 mM NaCl standard. Swirl the solution for 15 sec to help the ISE membrane to equilibrate with the solution, then leave undisturbed for a further minute. Note the multimeter reading (this is millivolts, mV) (the meter may fluctuate if you wave your hand over it so ensure that the environment is undisturbed when taking the reading).
  4. Remove, wipe and transfer the electrodes to the 20 mM standard and measure as before, waiting one minute before taking the reading.
  5. Repeat the procedure for the 40 mM and 80 mM standards.
  6. [Na+] (mM)

    10

    20

    40

    80

    Log10 [Na+]

           

    Meter reading (mV)

           

     

  7. Finally, take a reading in the same way from the undiluted redissolved sachet from Experiment A.
  8. Transfer the electrodes back to distilled water and, if nobody else is going to use the system, switch off the multimeter and amplifier (position O).
  9. Plot mV (y-axis) against log10 [Na+] (x-axis) and from this graph determine the apparent [Na+] in the redissolved sachet solution.

Meter reading for undiluted sachet solution

Log10 [Na+] from graph

[Na+] in redissolved sachet (mM)

 

 

   


2.5 Results:

  1. From the information provided on the sachet packaging, calculate the expected concentrations of Na+ and K+ ions in the solution you made.
  2. [Na+] (mM)

    [K+] (mM)

       

  3. How do your own values for Na+ and K+ determined by flame photometry and for Na+ determined with the ISE compare to the expected values?
  4. The class data for sodium determination only by both methods will be collected and tabulated on the blackboard in the class. Collect these data in the table below before you leave (12 sets maximum). Using a calculator, calculate the mean and standard deviation (S.D.) for the two data sets and compare these for accuracy and precision. What do you conclude?
  5.  

    1

    2

    3

    4

    5

    6

    7

    8

    9

    10

    11

    12

    [Na+](flame)

                           

    [Na+] (ISE)

                           

     

    Sodium concentration (mean ± S.D, n = ) by flame photometry = ……………… (mM)

     

    Sodium concentration (mean ± S.D, n = ) by ISE = ……………… (mM)

 

NOTE: If you do not have a nifty calculator that automatically calculates S.D.'s, or if you do not have time to collect the class data, an alternative method is provided for you for calculating the mean values ± S.D. for the class data and, if you are especially keen, for the whole year data (all 5 classes). The data sets from all 5 classes will be posted after each class on the "Class data page". An Excel spreadsheet can be downloaded from this page and the data entered following the simple instructions provided. The spreadsheet will automatically calculate and display the mean and S.D. for the data sets you enter. If you are not used to using a spreadsheet like this, this is a good opportunity to try a simple exercise. You will find such things very useful in the future.


2.5 Experiment C: Correct use of automatic pipettes

As there is sufficient free time during the course of this practical, you will take the opportunity to practice the correct use of the automatic pipettes that will be used extensively in later practical classes. It is essential that you learn how to use these properly if you are going to achieve reliable results in these classes. This exercise also provides another demonstration of accuracy and precision.

Accuracy: Assuming the pipette you use has been accurately calibrated (unfortunately not guaranteed), the closeness of the values you measure to the expected values will show how accurate your technique is. Any significant inaccuracy with a low standard deviation would indicate a consistent error in your technique, e.g. always (wrongly) pushing the pipette button fully down before drawing liquid into the pipette (or an improperly calibrated pipette).

Precision: Even though your mean value may appear accurate, if your individual values vary all over the place (high standard deviation), then your technique obviously suffers from a variety of errors (i.e. general carelessness!).


       Procedure:

  1. Read the instructions on the use of pipettes in Appendix 2and watch and listen carefully to the demonstration that will be given.
  2. Practice pipetting water out of and into a beaker, watching all the time what you are doing and checking the water in the tip for air spaces, drops of water left at the top of the tip after expelling etc. Show the demonstrator your technique before proceeding to the next step.
  3. Place a plastic weighing boat on the balance and tare the balance so that it reads zero (this will be demonstrated).
  4. Using a "blue button" pipette, pipette 1.0 ml (1000 µl) distilled water into the weighing boat and record the weight. If the pipette has been accurately calibrated, what should the reading be?
  5. Pipette another 1.0 ml water into the boat and record the weight again.
  6. Repeat this until you have taken 10 readings (10 ml total pipetted) then calculate the weight of each individual 1.0 ml sample.
  7. Calculate the mean and standard deviation of your 10 measurements.

1 ml samples

Sample

1

2

3

4

5

6

7

8

9

10

Weight (g)

                   





      Mean weight  = .................. g    Standard deviation =  ± .................. g



Return to .Practicals Home Page...... Return to .Practical 2: Main page...... Return to .Practical 2: Theory